71

Remote Interface Reference

System-Related Commands

5.3.3

Device Clear

A device clear can be initiated through the serial interface by sending the
<Ctrl-C> (ASCII code 03) character. The status registers and the error
queue are left unchanged when the device clear character is received.
Device clear performs the following actions:

The Model 286 serial output buffer is cleared.

The Model 286 is prepared to accept a new command.

5.4 Command Reference

The following paragraphs present all instrument commands and queries
in related groups and a detailed description of the function of each
command or query is provided. Examples are also provided where
appropriate. Return strings may be up to 80 characters in length.

5.4.1

System-Related Commands

*IDN?

Returns the instrument’s identification string. The identification string
contains the AMI model number and firmware revision code.

*RST

Resets the instrument. All non-volatile calibration and instrument
settings are restored. Status is cleared.

*TST?

Performs a self-test. Currently always returns “1” plus a diagnostic code
useful for troubleshooting by an Authorized AMI Technical Support
Representative.

<

Ctrl-C

>

This command clears the output buffers of the instrument and prepares
the instrument for a new command. Status registers are unaffected. <

Ctrl-

C

> corresponds to ASCII code 03.

SYSTem:LOCal

Enables all front panel controls. All front panels controls are enabled by
default after a power-up or

*RST

command.

SYSTem:REMote

Disables all front panel controls.