Operation

Shift Key Commands : Reset Quench

If Voltage Limit becomes active while ramping, it will be indicated by a

reverse illumination character “V” for the status indicator.

Once the Voltage Limit function becomes active, the current, and therefore

field, will no longer be ramping linearly with time as the voltage available

to charge the magnet will be reduced as the total loop voltage will be

limited. As the IR drop of the leads increased with current, the voltage

available to charge the magnet will be reduced.

3.7.3

Reset Quench SHIFT-key

The

RESET QUENCH

SHIFT-key is used whenever a quench detection has

occurred and is being indicated on the display (example shown below).

When a quench detection has occurred, the Model 430 Programmer will

respond to no further input until the

RESET QUENCH

SHIFT-key is used,

or until the quench condition is cleared by a remote command. See Refer to

section 3.15 on page 88.

3.7.4

Increment Field SHIFT-key

The

INCR. FIELD

SHIFT-key is used to manually increase the field. This

is done at the defined ramp rate.

When the

INCR. FIELD

SHIFT-key is used, the current/field begins

ramping up. If the

INCR. FIELD

SHIFT-key is used again (while the

current/field is manually ramping up), the ramping will be paused.

Alternately, the

RAMP / PAUSE

key may be pressed to pause manual

ramping. Manual ramping will continue until paused or the Current

Limit

1

or Magnet Current Rating is achieved.

3.7.5

Field <> Current SHIFT-key

The

FIELD <> CURRENT

SHIFT-key is used to toggle between the use of

field units, either kG (kilogauss) or T (tesla), and the use of current units

(A)

2

. If the Model 430 Programmer is using field units (either kG or T) and

1. Refer to section 3.10.2.4 on page 54.
2. The value is always displayed in current (A) when an installed persistent switch is in

the

cooled state

since the value represents power supply current only, independent

of magnet current/field.

+40.92 A

Mode: Ramping

+2.50 Vs

V

PSwitch: ON

+0.00 A - Quench Detect @ +45.81 A
+0.00 Vs

PSwitch Heater: ON