6 decrement field shift-key, 7 field units shift-key, 8 persistent switch heater current shift-key – American Magnetics 05200PS-430-601 Integrated Power Supply System User Manual

Page 58: Operation

Advertising
background image

40

Rev. 5

Operation

Shift Key Commands : Decrement Field

the

FIELD <> CURRENT

SHIFT-key is used, the Model 430 Programmer

will begin using current units (A). Conversely, if the Model 430 is using

current units (A) and the

FIELD <> CURRENT

SHIFT-key is used, the

Model 430 will begin using field units (either kG or T).

Note

The Model 430 Programmer cannot use field units unless a valid

coil constant has been entered

1

.

3.7.6

Decrement Field SHIFT-key

The

DECR. FIELD

SHIFT-key is used to manually decrease the current/

field. This is done at the defined ramp rate.

When the

DECR. FIELD

SHIFT-key is used, the current/field begins

ramping down. If the

DECR. FIELD

SHIFT-key is used again (while the

current/field is manually ramping down), the ramping will be paused.

Alternately, the

RAMP / PAUSE

key may be pressed to pause manual

ramping. Manual ramping will continue until paused or the Current

Limit

2

or Magnet Current Rating is achieved.

3.7.7 Field

Units

SHIFT-key

Use of the

FIELD UNITS

SHIFT-key provides a shortcut to the picklist

menu

3

for defining whether the field is specified and displayed in units of

kilogauss (kG) or tesla (T). The selected option also applies to remote

interface commands. The default setting is kilogauss.

3.7.8

Persistent Switch Heater Current SHIFT-key

Use of the

P. SWITCH HTR. CUR.

SHIFT-key provides a shortcut to the

menu

4

for setting persistent switch heater current. The value can be set to

between 0.0 and 125.0 mA. The default value is 10.0 mA unless preset by

AMI to match a specific superconducting magnet.

1. Refer to section 3.10.2.2 on page 52.
2. Refer to section 3.10.2.4 on page 54.
3. Refer to section 3.10.3.4 on page 62.
4. Refer to section 3.10.2.8 on page 57.

+50.00 A - Field Units

+0.50 Vs

Kilogauss

Tesla

+50.00 A — PSwitch Current (mA)

+0.50 Vs

10.0

Advertising