Operation

Shift Key Commands : Stability

3.7.9 Stability

SHIFT-key

Use of the

STAB.

SHIFT-key provides a shortcut to the menu for defining

the Model 430 stability setting. The stability setting is specified in percent

and controls the transient response and stability of the system. The valid

input range is from 0.0 to 100.0%. The default value is 0.0% unless preset

by AMI to match a specific superconducting magnet.

See section 3.10.2.1 on page 50 for details of how to determine the stability

setting to use.

3.7.10 Vs <> Vm SHIFT-key

The

Vs <> Vm

SHIFT-key is used to toggle the voltage display between

display of the voltage across the magnet (Vm) and the power supply output

voltage (Vs). See section 3.2.2 on page 26 for details.

3.7.11 Volt

Meter

SHIFT-key

The

VOLT METER

SHIFT-key is used to toggle the main display between

display of a voltmeter indicating magnet voltage (Vm) or supply voltage

(Vs), and display of ramp mode and persistent switch heater state. See

section 3.2.4 on page 27 for details.

3.7.12 Fine

Adjust

SHIFT-key

The

FINE ADJUST

SHIFT-key is used to enable the use of the front panel

fine adjust knob to adjust numeric values. See section 3.4 on page 28 for

details.

3.7.13 Persist. Switch Control SHIFT-key

Refer to section 3.6.1 on page 31.

3.8

LED Indicators

The Model 430 Programmer has six front panel LED indicators. See figure

with Table 1-1 on page 5 for the location of these indicators.

3.8.1 Power-on

Indicator

The green power-on LED indicates that the Model 430 Programmer is

powered on.