18

Rev.

3

Installation

Power-Up Procedure

doubt as to the correct connection of a component, contact an AMI

Technical Support Representative. The user may be required to

properly make a few connections between the various system

components which were disconnected to facilitate packing and

shipping.

2. Temporarily place a short across the magnet current terminals.

Often this is most easily accomplished by unfastening the heavy

cables from the magnet current leads and fastening them together.

1

This will allow rudimentary power supply checks without

energizing the superconducting magnet.

3. Energize the Model 430 Programmer by placing the power switch

in the

I (ON) position.

4. When prompted by the Model 430 Programmer, energize the power

supply and press ENTER on the Model 430 Programmer.

Note

The slave unit must be turned on between 2 to 5 seconds after the

master, otherwise the units will fault and the power-up for bothall

units will have to be repeated.

Warning

All power supply parameters, both hardware and software, have

been set at the AMI factory. Power supply control, with the exception

of powering

ON

and

OFF

sequencing, is done by way of the AMI

Model 430 Power Supply Programmer. No field adjustments or

reconfiguration of the power supply should be attempted in the field

unless specifically described in this document or recommended in

writing by an AMI Technical Support Representative.

5. Enter a stability setting of 100%. Refer to sections 3.3 on page 25

and 3.7.9 on page 39.

6. Verify the various setup menu values for the system (with the

exception of the stability setting, which is to be temporarily left at

100%). If the power supply system was purchased with an AMI

magnet, AMI has preset the setup menu values for proper

operation. See sections 3.3, 3.5, 3.9 and 3.10 for more discussion of

the setup menu values and their entry into the Model 430

Programmer.

1. If the system shipped with CamLoc quick-disconnect connectors, they may be quickly dis-

connected from the magnet leads and connected together.