Rev. 3

19

Installation

Power-Up Procedure

7. Set the Model 430 Programmer to display current (rather than

field). Refer to sections 3.2.1 and 3.7.5.

8. Set the ramp rate to 1 A/sec. Refer to sections 3.3 on page 25 and

3.7.1 on page 33.

9. Set the target current to 10 A. Refer to sections 3.3 on page 25 and

3.6.2 on page 31.

10. If a Persistent Switch is installed, set the PSw P/S Ramp Rate to 10

A/sec. Refer to paragraph 3.10.2.11 on page 56

11. Initiate ramping to the target current by pressing the

RAMP /

PAUSE

key (status indicator changes from

P

to ).

12. The system should ramp to 10 A in approximately 10 seconds.

Verify this is the case (if a PSwitch is installed and in the cooled

state, ramp time to 10 A should be slightly less than 2 seconds).

13. When the target current is achieved, the

FIELD AT TARGET

LED

will be illuminated. The display should show “

+10.00 A -

indicating that the Model 430 Programmer is in the holding mode

at the target current value (+10.00 A).

Note

There may be a discrepancy between the current shown on the power

supply display

1

and the current displayed on the Model 430

Programmer. The current measurement system incorporated in the

Model 430 is more accurate than the power supply shunt.

14. Verify that the output current display of the power supply indicates

that it is supplying 10 A to the load (which is only the cabling in

this case).

15. Set the target current to the master current limit value. Refer to

section 3.10.2.4 on page 52 to determine the master current limit

value. After the new target current value is entered, the Model 430

Programmer should ramp automatically to the new setting.

16. When the new target current value is reached, the power supply

current display (if provided) should also indicate the new value.

17. Press the

RAMP TO ZERO

key to ramp the system to zero current.

18. Perform remote control software checkout as required.

1. Not all power supplies have a local current readout.