Ec control board, Figure 15, Ec30 control board – M&C TechGroup EC-30_FD Operator's manual User Manual
Page 31: 20 ec control board, Channel 2 channel 1
Advertising

31
Gas sampling and gas conditioning technology
3-4.20-ME
20
EC CONTROL BOARD
Figure 15 shows the layout of the EC30 control board (wiring diagram in appendix).
Figure 15 EC30 control board
Temperature
adjustment
(zero)
Alarm threshold
(<-25°C at MP3)
Power
Trans-
former
Solenoid
valve
Power
supply
Channel 2
Channel 1
Channel
Operating indication
Heating
Operati
o
n
T
es
t
Voltage at MP1, MP2, MP3 measured
against MP5 (V):
1
2
3
4
5
6
7
8
Corresponding temperatures in °C
+30
+20
+10
0
-10
-20
-30
-40
Voltage at MP1, MP2, MP3 measured
against MP5 (V):
1
2
3
4
5
6
7
8
Corresponding temperatures in °C
+30
+20
+10
0
-10
-20
-30
-40