Gas sampling and gas conditioning technology
3-4.20-ME
The sample gas flows through the pre-cooling stage
and low-temperature stage . This cycle will
be changed automatically every three hours.
In the event of a brief (mains) power failure:
The current cycle status will be stored;
All LEDs go out;
The status alarm contact is activated;
The flow of sample gas is cut off externally, in case the sample gas pump or a solenoid valve in
the sample gas line are to be controlled by an alarm contact.
When the mains power is restored:
The controller starts up automatically;
The status alarm contact is deactivated;
The status alarm contact will release the flow of sample gas externally, in case the sample gas
pump or a solenoid valve in the sample gas line are controlled by an alarm contact.
In the event of a prolonged power failure, the controller will start up automatically like it happens for a
new start. Release is only done when the temperature is < -25°C.