M&C TechGroup SP2400-H Operator's manual User Manual
Page 19

2-1.1-ME
Gas sampling and gas conditioning technology
19
N O T E !
A preferred mounting position is to have the probe with its sample gas outlet
pointing downwards, although this is not absolutely necessary for perfect
functioning.
Insert flange seal (Fig. 6) between sampling flange and probe flange.
If the heated probe tube type SP30/35 or the ceramic prefilter type V12 is used, then the probe is to
be screwed to their flange (Fig. 6)(with welded threaded bolt). First insert the flange seal be-
tween the two flanges.
If the probe connection does not correspond to the standard flange connection DN65 PN6, then the
optionally supplied adapter flange (Fig. 2 and 6) should be mounted to the probe in the same way.
For high pressure version /HP a flange ist standardmäßig ein Flansch DN50 PN25 vorhanden
Screw either the probe tube or prefilter (Fig. 6) with thread G3/4”o directly or using an extension
tube with the ¾" flat gasket into the ¾" inner thread of the flange of the probe and tighten.
Insert sample tube or prefilter of the complete probe unit into the connection piece and screw
probe at the connection piece using the screws, spring rings and nuts supplied.