14.4
MOUNTING OF SAMPLE LINE
Open sample line mounting bracket .
Push end of the sample line into silicone cap in the bottom of the angle bracket.
Depending on type of heated sample line insert stainless steel pipe end with or without PTFE line
through hole in the silicone cap .
Figure 7
Mounting of heated sample line
Connect stainless steel connection piece or interchangeable PTFE tube at the fitting in the sam-
ple gas outlet of the probe. For this remove union nut with ferrule and cutting ring and put it in right
order and direction on the stainless steel connection piece or the PTFE tube.
A T T E N T I O N !
If a PTFE tube is used as sample line, an insert must under all circum-
stances be inserted in the end of the tube in order to prevent the tube be-
ing pressed together.
Put connection piece or tube end into the fitting in the sample gas outlet of the probe and tighten
the union nut by hand.