5. MAINTENANCE

Caution: before cleaning, turn off the scale, and remove the AC adapter.

5.1 Cleaning

The housing may be cleaned with a cloth dampened with a mild detergent if
necessary.
Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the
housing or control panels.

5.2 Troubleshooting

The following table lists common problems and possible causes and remedies. If
the problem persists, contact Ohaus or your authorized dealer.

TABLE 5.1. TROUBLESHOOTING.

Symptom Possible

Cause(s)

Remedy

Scale will not turn on No power to scale

Battery power used up

Verify connections and voltage.
Verify battery polarity

Poor accuracy

Improper calibration. Unstable
environment

Perform calibration. Move scale to
suitable location

Unable to calibrate

CAL LOCK set to ON
Unstable environment
Incorrect calibration mass

Set CAL LOCK to OFF (see section
4.2)
Move the scale to suitable location
Use correct calibration mass

Scale displays “Lo
bAt”

Battery is discharged

Connect power or change the
batteries

Scale displays “Lo
Line”

Low Line voltage

Verify source voltage

Scale displays “E”

Weight on pan exceeds
capacity

Remove weight from the pan

Scale displays “CALE” Incorrect Calbiration weight on

pan

Use correct calibration weight

Scale displays “UnSt” Scale is unstable at power up. Relocate scale to a more stable

environment