PREPARATION

The Navigator Balance has two calibration processes: Span and Linearity.
Follow the prompts for calibration mass values.

Note: This menu may be locked, either for security or for Legal for Trade applications. To regain
access, see Instruction Manual, Section 4.6. (For more detail, see Service Manual, Section 1.7.)

 Ensure that the balance is level, stable, and undisturbed during calibration.

(Otherwise the process could fail.)

 Allow balance to warm up for five minutes after stabilizing to room temperature.

 Use suitable calibration masses (available from Ohaus).

 Before beginning Calibration, be sure the

“lock switch” is off.

(See note above.)

After calibration, the balance returns to weigh mode.

Note: To cancel calibration, long-press:

SPAN CALIBRATION

5

(Release button)

Y

1

(Acquiring zero value)

Y

1

(Place mass on platform)

Y

1

(example for 200g capacity model)

5

= Long Press

1

= Short Press