2 phase check, 3 pi index check – Soft dB Opus Suite Intensity Module User Manual

Page 13

Advertising
background image

11

5.1.2

Phase Check

To perform phase check:

1. Connect the intensity calibrator to the output 1 of the Concerto.

2. Insert both microphones calibrator cavity.

3. Click on the Phase Check button on the Calibration Process interface to launch the Phase

Check interface.

4. Click START to begin the test. An averaging is performed on the signal and the resulting phase

response is displayed on the interface. An acceptable tolerance is also displayed on the graph.

5. When averaging is done, click OK to accept.

5.1.3

PI Index Check

To perform PI Index check:

1. Perform Amplitude calibration and Phase check priorly.

2. Connect the intensity calibrator to the output 1 of the Concerto.

3. Insert both microphones calibrator cavity.

4. Click on the PI Index Check button on the Calibration Process interface to launch the PI Index

Check interface.

5. Click START to begin the test. An averaging is performed on the signal and the resulting PI

Index response is displayed on the interface. A minimal value is also displayed on the graph.

6. When averaging is done, click OK to accept.