English

11

PCS10/K8047 © 2003 Velleman Components

Save data :

Saves the waveform in text format. only the portion of the data displayed on the

screen is saved.

AutoSave data :

Saves the image and its

date

to a file during the run of a sample.

Print :

Print the image.

Print setup :

Selects a printer and sets printer options before printing. The available options

depend on the printer you select.

Exit :

Teminates the program.

1.5.1.1

Date

Start : start time of the recording

Time step : Timescale setting 100 samples = 1second

Voltage range CH1 : Voltage range channel 1 => measured value 255 corresponds to 15V
CH2 : Voltage range channel 2 => measured value 255 corresponds to 30V
CH3 : Voltage range channel 3 => measured value 255 corresponds to 30V
CH4 : Voltage range channel 4 => measured value 255 corresponds to 30V

1700 measurements values are taken per display, numbered from 0 to 1700. In addition to this, the
measured values can be read from channel 1 to channel 4.

Example CH1 :

Point in time : 13 x 1/100 = 0,13s + 9:17:02 = 9:17:02.13
Voltage : 138 x 15/255 = 8,118V

Example CH3 :

Point in time : 4 x 1/100 = 0,04s + 9:17:02 = 9:17:02.04
Voltage : 61 x 30/255 = 7,176V