VEGA VEGAPULS 68 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA_HART four-wire User Manual

Page 24

Advertising
background image

be a place where no installations cross the microwave signals. Make
sure when planning your measurement loop that the radar signals
have a "clear view" to the product.

I

n case of existing vessel installations, a false echo storage should be

carried out during setup.

I

f large vessel installations such as struts or supports cause false

echoes, these can be attenuated through supplementary measures.
S

mall, inclined sheet metal baffles above the installations scatter the

radar signals and prevent direct interfering reflections.

Fig. 15: Cover smooth profiles with deflectors

T

o avoid strong buildup and dust in the antenna system, the sensor

should not be mounted directly at the dust extraction of the vessel.

Tip:
I

n case of extreme dust in the antenna system, VEGAPULS 68 is

available with rinsing connection e.g. for air. The air is distributed via
channels in the antenna system and keeps it virtually free of dust.

Fig. 16: Purging air connection with horn antenna

Air purging

24

VEGAPULS

68 • 4 … 20

mA/HART four-wire

4 M

ounting

29262

-EN

-090305

Advertising