VEGA VEGADIF 65 4 … 20 mA_HART User Manual

Page 54

Advertising
background image

54

6 Adjustment with the display and adjustment module PLICSCOM

VEGADIF 65 • 4 … 20 mA/HART

36128-EN-130417

Menu section

Menu item

Reset value

Basic settings

Zero/Min. adjustment

Measuring range begin

Span/Max. adjustment

Measuring range end

Density

1 kg/l

Density unit

kg/l

Damping

1 s

Linearization

Linear

Sensor-TAG

Sensor

Display

Displayed value

Differential pressure

Display unit

Mass/kg

Scaling

0.00 to 100.0

Decimal point indication 8888.8

Diagnostics

Totalizer

0.0000 10

00

gal

Part sum counter

0.0000 10

00

gal

Service

Current output - charac-

teristics

4 … 20 mA

Current output - failure

mode

< 3.6 mA

Current output - min.

current

3.8 mA

Current output - max.

current

20.5 mA

The values of the following menu items are not reset with "Reset:

Menu section

Menu item

Reset value

Basic settings

Unit of measurement

bar

Temperature unit

°C

Position correction

No reset

Display

Backlight

No reset

Service

Language

No reset

HART mode

No reset

Applications

No reset

Peak value

The min. and max. temperature or pressure values are each reset to

the actual value.
Totalizer

The total and part sum counter are reset to zero.

Additional adjustment and diagnosis options such as e.g. scaling,

simulation or trend curve presentation are shown in the following

menu schematic. You will find a detailed description of these menu

items in the operating instructions manual "Display and adjustment

module".

Optional settings

Advertising