6 Adjustment with the display and adjustment module PLICSCOM

VEGADIF 65 • 4 … 20 mA/HART

36128-EN-130417

6.5 Menu schematic

Information:

Depending on the version and application, the highlighted menu

windows may not always be available.

Basic adjustment differential pressure

Basic adjustment

1

Display
Diagnostics
Service
Info

Applications

1.1

Differential pressure

Unit

1.1

Unit of measurement
bar

Temperature unit
°C

Position correction

1.2

Offset

DP

=
-0.0035 bar
0.0000 bar

Zero

1.3

000.0 %

DP

=
0.0000 bar
0.0000 bar

Span

1.4

100.00 %

DP

=
0.5000 bar
0.0000 bar

Damping

1.5

1 s

Linearization curve

1.6

Linear
Horizontal cylindrical tank
Spherical tank
User programmable

Sensor-TAG

1.7

Sensor

Basic setting level

Basic adjustment

1

Display
Diagnostics
Service
Info

Applications

1.1

Level

Unit

1.1

Unit of measurement
bar

Temperature unit
°C

Position correction

1.2

Offset

DP

=
-0.0035 bar
0.0000 bar

Min. adjustment

1.3

000.0 %
=
0.0000 bar
0.0000 bar

Max. adjustment

1.4

100.00 %
=
0.5000 bar
0.0000 bar

Damping

1.5

1 s

Linearization curve

1.6

Linear
Horizontal cylindrical tank
Spherical tank
User programmable

Sensor-TAG

1.7

Sensor