4 Mounting

VEGAPULS 62 • Modbus and Levelmaster protocol

41362-EN-121011

Instructions for orientation:

•

Note marking of the polarisation plane on the sensor

•

With threaded fitting, the marking is on the hexagon, with flange

connection between the two flange holes

•

The marking must show to the direction of the tube connections to

the vessel

Instructions for the measurement:

•

The 100 % point must not be above the upper tube connection to

the vessel

•

The 0 % point must not be below the lower tube connection to the

vessel

•

Min. distance sensor reference plane to the upper edge upper

tube connection > 300 mm

•

For the parameter adjustment, select "Application standpipe" and

enter the tube diameter to compensate errors due to running time

shift

•

A false signal suppression with integrated sensor is recommended

but not mandatory

•

The measurement through a ball valve with complete run is pos-

sible

Constructional requirements on the bypass pipe:

•

Material metal, smoother inner tube

•

In case of an extremely rough tube inner surface, use an inserted

tube (tube in tube) or a radar sensor with tube antenna

•

Flanges are welded to the tube according to the orientation of the

polarisation level

•

Gap size with junctions ≤ 0.1 mm, for example, when using a ball

valve or intermediate flanges with single pipe sections

•

The antenna diameter of the sensor should correspond to the

inner diameter of the tube

•

Diameter should be constant over the complete length

Instruments for a temperature range up to 250 °C or up to 450 °C

have a distance piece between process fitting and electronics hous-

ing. Ths distance piece is used for thermal decoupling of the electron-

ics against high process temperatures.
Information:

The distance piece must only be incorporaed up to max. 50 mm in

the vessel installation. Only then, a reliable temperature decoupling is

guaranteed.