4 Mounting

VEGAPULS 62 • Modbus and Levelmaster protocol

41362-EN-121011

3 ... 4 h

max

≥ 50 mm

90°

2

3

1

h

ma

x

d

mi

n

≥ 2 mm x

h

ma

x

90°

4

2

3

Fig. 24: Flow measurement with rectangular flume: d

min.

= min. distance of the

sensor (see chapter "Technical data"); h

max.

= max. filling of the rectangular

flume
1 Overflow orifice (side view)

2 Headwater

3 Tail water

4 Overfall orifice (view from bottom water)

In general, the following points must be observed:

•

Install the sensor on the headwater side

•

Installation in the centre of the flume and vertical to the liquid

surface

•

Distance to the overfall orifice

•

Distance of orifice opening above ground

•

Min. distance of the orifice opening to bottom water

•

Min. distance of the sensor to max. storage level