VEGA VEGABAR 82 Profibus PA - Operating Instructions User Manual

Page 68

Advertising
background image

68

11 Supplement

VEGABAR 82 • Profibus PA

45031-EN-131107

Ambient conditions

Version

Ambient temperature

Storage and transport temper-

ature

Standard version

-40 … +80 °C (-40 … +176 °F)

-60 … +80 °C (-76 … +176 °F)

Version IP 66/IP 68 (1 bar)

-20 … +80 °C (-4 … +176 °F)

-20 … +80 °C (-4 … +176 °F)

Version IP 68 (25 bar) with connec-

tion cable PUR

-20 … +80 °C (-4 … +176 °F)

-20 … +80 °C (-4 … +176 °F)

Version IP 68 (25 bar), connection

cable PE

-20 … +60 °C (-4 … +140 °F)

-20 … +60 °C (-4 … +140 °F)

Process conditions
Product temperature depending on the measuring cell seal

2)

The specifications are an overview. The specification on the type plate are valid.

Measuring cell seal

Product temperature - standard

version

Product temperature - version

with extended temperature range

FKM (VP2/A)

-20 … +130 °C (-4 … +266 °F)

-20 … +150 °C (-4 … +302 °F)

FKM (A+P 70.16)

-40 … +130 °C (-40 … +266 °F)

-

FKM (ET 7067)

-20 … +130 °C (-4 … +266 °F)
1 h: 140 °C/284 °F cleaning tem-

perature

-

EPDM (A+P 75.5/KW75F)

-40 … +130 °C (-40 … +266 °F)
1 h: 140 °C/284 °F cleaning tem-

perature

-40 … +150 °C (-40 … +302 °F)

EPDM (ET 7056)

-40 … +130 °C (-40 … +266 °F)
1 h: 140 °C/284 °F cleaning tem-

perature

-

FFKM (Kalrez 6375)

-20 … +130 °C (-4 … +266 °F)

-20 … +150 °C (-4 … +302 °F)

FFKM (Perlast G75S)

-15 … +130 °C (-4 … +266 °F)

-15 … +150 °C (5 … +302 °F)

FFKM (Perlast G75B)

-15 … +130 °C (-4 … +266 °F)

-15 … +150 °C (5 … +302 °F)

FFKM (Chemraz 535)

-30 … +130 °C (-22 … +266 °F)

-

FEPM (Fluoraz SD890)

-5 … +130 °C (-22 … +266 °F)

-

Product temperature depending on the ambient temperature

Ambient temperature

Product temperature

+50 °C (+122 °F)

+150 °C (+302 °F)

+80 °C (+176 °F)

+100 °C (+212 °F)

Mechanical stress, depending on the instrument version
Vibration resistance

4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration

with resonance)

Shock resistance

100 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

2)

With process fitting PVDF, max. 100 °C (212 °F).

Advertising