• Never use too much water.

• Do not bend the cuff or sleeve. Wrap the inside to the outside.

• Do not dismantle the unit, remove the cuff or try to repair yourself. Contact your distributor.

• Do not operate the unit under severe climatic circumstances or at extreme temperatures,

humidity or direct sunlight.

18. Troubleshooting and Error Messages


Trouble/Error Possible

Cause

Solution

No indication when powered

on

• Batteries are not properly

installed

• Drained batteries

• Check the polarity of the

batteries

• Remove the batteries and

wait 10 seconds before

reinstalling

Err 0

• No pulse or detected pulse

too low

• Refit the cuff and repeat

the measurement

Err 1

• Cuff is leaking air or

inflation is too slow

• Replace the batteries or

contact your distributor

Err 2

• Cannot measure systolic

• Refit the cuff and repeat

the measurement

Err 3

• Cuff pressure is too high

• Refit the cuff and repeat

the measurement

Err 4

• Cannot measure systolic

or is outside of range

• Refit the cuff and repeat

the measurement

Err 5

• Deflation is too fast

• Contact your distributor

Err 6

• Deflation is too fast

• Contact your distributor

Err 7

• Systolic is outside of range

• Refit the cuff and repeat

the measurement

Err 8

• Main pressure is outside of

range

• Refit the cuff and repeat

the measurement