b. Fuse Replacement


• Set the rotary switch to the OFF position.

• Disconnect all test leads and/or any connectors from the terminals.

• Open the battery compartment using an appropriate screwdriver.

• Remove the fuse by sliding it out of its bracket.

• Install a replacement fuse (F 500mA/250V, Ø 5 x 20mm).

• Close the battery compartment.

c. Battery Replacement

• Set the rotary switch to the OFF position.

• Disconnect all test leads and/or any connectors from the terminals.

• Open the battery compartment using an appropriate screwdriver.

• Remove the battery.

• Install a replacement battery (6F22).

• Close the battery compartment.

6. Technical Specifications


Accuracy is specified for one year after calibration at operating temperatures of 18°C ~ 28°C with a relative humidity

at 0% ~ 75%. Accuracy specifications take the form of ± (% of rdg + number of least significant digits).

Environmental Conditions

1000V CAT. II and 600V CAT. III

Pollution Degree

2

Altitude <

2000m

Operating Temperature

0°C~40°C or 32°F~122°F (< 80% RH, < 10°C)

Storage Temperature

-10°C~50°C or 14°F~122°F (< 70% RH, battery removed)

Temperature Coefficient

0.1x / C° (< 18°C or > 28°C)

Max. Voltage between Terminals and Earth 600VAC RMS or 600VDC
Fuse Protection

F500mA/250V, 5 x 20mm quick fuse

Sample Rate

2x/sec for digital data

Display 3

5/6

digits LCD with automatic indication of functions and symbols

Over Range Indication

yes (“OL”)

Polarity Indication

“-“ displayed automatically

Auto Power-Off

yes, after 15 min

Continuity Buzzer

yes

Transistor Test

no

Diode Test

yes, 1.0mA (open circuit voltage = 1.5VDC)

Low Battery Indication

yes ( )

Max. Display

3,999 counts

Ranging Modes

manual or automatic

Data Hold

yes

Backlight yes
Power Supply

9V battery

Dimensions

180 x 85 x 45mm

Weight

± 280g (with battery)

Accessories

1 x K-type thermocouple, 2 x test leads, 1 x 9V battery, manual

Before fuse replacement, disconnect all test leads and/or ant connectors from any circuit

under test. To avoid damage to the instrument, replace the fuse only with specified ratings.

To avoid false readings, which could lead to possible electric shocks or personal injury,

replace the battery as soon as the battery indicator appears. Before battery replacement,

disconnect all test leads and/or ant connectors from any circuit under test.