Velleman EVM1400 User Manual

Page 9

Advertising
background image

EVM1400_v2

PEREL

9

DC Voltage

Range Resolution Accuracy

400mV 0.1mV

± (0.7% of rdg + 2 digits)

4V 1mV

40V 10mV

400V 100mV
600V

1V

± (1.0% of rdg + 2 digits)

Overload protection: 250VDC or 250VAC RMS for 400mV range; 600VDC or 600VAC RMS for 4V~600V ranges
Input impedance: 10MΩ

Max. input voltage: 600VDC

Note that the meter may display an unsteady reading when the test leads do not touch the circuit due to the high
sensitivity.


AC Voltage

Range Resolution Accuracy

4V 1mV

± (0.8% of rdg + 3 digits)

40V 10mV

400V 100mV
600V

1V

± (1.5% of rdg + 5 digits)

Overload protection: 600VDC or 600VAC RMS for 4V~600V ranges

Input impedance: 10MΩ
Max. input voltage: 600VAC

Frequency range: 40~400Hz
Response: avg, calibrated in RMS of the sine wave


Note that the meter may display an unsteady reading when the test leads do not touch the circuit due to the high

sensitivity.

DC Current

Range Resolution Accuracy

40mA

10µA

± (1.2% of rdg + 3 digits)

400mA

100µA

± (1.2% of rdg + 3 digits)

10A

10mA

± (2.0% of rdg + 10 digits)

Overload protection: F500mA/250V quick fuse for mA ranges; unfused for 10A range

Max. input current: 400mA for mA terminal; 10A for 10A terminal
Voltage drop: 5mV/1mA for mA ranges; 10mV/1A for 10A range


AC Current

Range Resolution Accuracy

40mA

10µA

± (1.5% of rdg + 5 digits)

400mA

100µA

± (1.5% of rdg + 5 digits)

10A

10mA

± (3.0% of rdg + 10 digits)

Overload protection: F500mA/250V quick fuse for mA ranges; unfused for 10A range

Max. input current: 400mA for mA terminal; 10A for 10A terminal
Voltage drop: 5mV/1mA for mA ranges; 10mV/1A for 10A range

Frequency range: 40~400Hz

Response: avg, calibrated in RMS of the sine wave