DVM173SD

16/12/2010

©Velleman nv

4

File (F)

Save…

save the recorded data to disk

Open…

open a saved file

Print Setup…

change the printer and printing options

Print Preview

print preview the graph or list

Print…

print the graph or the list

Instrument (I) Calibration Adjustment… calibrate the data logger to eliminate measuring errors

e.g. die to aging of the microphone

Datalogger Setup…

set up sampling rate, alarm levels, LED indications,
recording mode…

Download Data…

download data from the data logger

Stop Running…

stop reading data from the data logger (only in real-
time mode)

Previous Measurement

check the data from last measurement

Window (W)

Instrument

show or hide the instrument window (left pane)

Infor Dialog

show or hide info dialog window

Help (H)

About

show the software version of the data logger

Help Topics

open the software help file

Open file save dialog box to save a data file. The file is automatically saved with a .sound
extension as well as with a .xls extension.

Open previously saved files (.sound extension).

Calibrate the data logger. E.g. when reading is 0.5dB higher than actual sound level, set the
adjustment value to -0.5dB.

Data logger setup.

• Set the sampling rate, sampling points, alarm levels (Hi and Low) and configure the LED

indications.

• Select normal (NORM) measurement or peak values (PEAK) only.

• Select storing data (STOR) or real-time (RT) measurement.

• Select the weighting type dBA (related to human hearing, with low and high frequency cut

off) or dBC (actual loudness, better for peak measurement).

• The logger speed can be set to FAST or SLOW.