3B Scientific Stereo Microscope, 40x, Transmitted-Light Illumination LED (230 V, 50__60 Hz) User Manual

3b scientific, Physics

Advertising
background image


3B SCIENTIFIC

®

PHYSICS



1

Stereo Microscope, 40x, Transmitted-Light Illumination, LED

115 V, 50/60 Hz: 1013369 / 230 V, 50/60 Hz: 1013128

Instruction Manual

07/13 ALF

1 Eyepiece

with

eye-shields

2 Tube

3 Lock-screw

4

Adjustment knob for focusing

5 Top-light

housing

6 Pillar

7

Toggle switch for illumination

8 Specimen

clips

9 Stand

10 Object

plate

11 Mains

switch

12 Objective

1. Description, technical data

The stereo microscope allows three-dimensional
viewing of objects in 20x and 40x magnification.

The microscope 1013369 is for operation with a
mains voltage of 115 V (±10%), and the
1013128 unit is for operation with 230 V (±10%).

Stand: Metal stand, column firmly connected
with base, pinion knobs attached on both sides
of the stand for coarse and fine focusing

Tube: Binocular inclined 45°, interocular dis-
tance adjustable between 55 and 75 mm, head
rotatable by 360°

Eyepieces: Pair of wide field eyepieces WF 10x
20 mm with eyepiece lock and rubber eyepiece
cups, diopter compensation ±5 on the left eyepiece

Objectives: Revolving nosepiece with objective 2x / 4x

Enlargement: 20x / 40x

Object Plate: Base with detachable object
plates (plastic, black/white and glass) 95 mm
diam. and 2 specimen clips

Illumination: LED, top and transmitted light
illumination, toggle switch to select light combi-
nation, power supplied by rechargeable battery,
6 V DC, 500 mA, battery charger 115 V or 230 V

Power supply: 1013369: 115 V, 50/60 Hz
1013128: 230 V, 50/60 Hz

Dimensions: 190 x 300 x 115 mm³ approx.

Weight: 2.9 kg approx.


2. Unpacking and assembly

The microscope is packed in a molded styro-
foam container.

Take the container out of the carton remove
the tape and carefully lift the top half off the
container. Be careful not to let the optical
items (objectives and eyepieces) drop down.

To avoid condensation on the optical compo-
nents, leave the microscope in the original pack-
ing to allow it to adjust to room temperature.