Rs232 default settings – Ocean Optics NeoFox Engineering Note User Manual
Page 26

NeoFox Communication Interfaces
26 013-20000-010-04-201211
Target
Baud
Actual
Baud
Relative
Error
Divisor
Latch
Multiply
Value
Divisor
Add Value
110
110
0.00%
6250
1
11
300
300
0.00%
2500
0
2
1200
1200
0.00%
625
0
2
2400
2400
0.00%
250
1
4
4800
4800
0.00%
125
1
4
9600
9603
0.03%
71
1
10
19200
19182
-0.10%
23
7
10
38400
38352
-0.12%
11
7
9
57600
57692
0.16%
13
0
2
115200
115385
0.16%
6
1
12
230400
230769
0.16%
3
1
12
460800
461538
0.16%
1
5
8
750000
750000
0.00%
1
0
2
A calculator is available at the following web address to help determine parameter values for baud rates that are
not listed in this table. You will need to use the value of 12,000,000 for the UART lock rate.
RS232 Default Settings
Setting
Value
Baud Rate:
57,600 baud (default)
Stop Bits:
1
Parity:
None
Flow Control:
None
Byte Order:
Little endian
Integer Size:
4 bytes
Float Size:
4 bytes