Ocean Optics NeoFox Engineering Note User Manual
Page 30
 
NeoFox Communication Interfaces
30 013-20000-010-04-201211
Single Point Temperature
float
X
≤ 200
188
 
Temperature (in degrees C) to use for next SPR calculation. Most users will want to set this to the current value of 
Temperature read by the instrument while in the SPR environment. 
 
Single Point Tau 
float
X ≤ 10.0
186
 
Tau (in microseconds) to use for next SPR calculation. Most users will want to set this to the current value of Tau 
read by the instrument while in the SPR environment. 
Single Point Oxygen
float
0 ≤ X
187
 
Percent Oxygen (in percent of 1 ATM) to use for next SPR calculation. Most users will want to set this to the known 
value of percent oxygen in the SPR environment. In ambient air, at sea level, this will be equal to 20.9. 
Single Point Calculate
N/A
N/A
189
 
Writing any value to this parameter will result in a single point reset with the single point reset parameter values 
listed above. The new single point coefficients will be calculated and the device will be put into single point mode 
(as opposed to multipoint mode). The new value will not necessarily be set in flash memory until the user manually 
issues a flash write command.