3: Software Operation
RAYSPHERE-1700-02-201304
13
Make the desired setting changes.
Setting
Meaning
*Delay of QE/NQ
Set the delay time for the QE65000 (QE) and/or NIRQuest (NQ) spectrometer in
the RaySphere. The available range is 0 – 160 ms.
Amplification
Sets the amplification of the photocurrent in two steps:
• Gain 1, Gain 2, Gain 3: amplify by a factor of 1, 10, 100, respectively
• Fine amplification by a factor of 1:10
The resulting overall amplification is: Amplification = GainX x fine amplification.
[GainX = {1; 2; 3}]
*Trigger Level
Adjusts the trigger level values for the photocurrent from 500 to 4094.
Trigger Source
Changes the trigger source:
• External – RaySphere reacts to an electrical trigger pulse which can be
applied on the power supply
• Flash – RaySphere triggers itself by flash if the trigger level is reached
• Manual – Spectrum recording is done manually. RaySphere measures
when the sample button on the Spectrometer page is pressed.
• Off – no triggering