9: Nonlinearity Correction, Stray Light Correction and Calibration
RAYSPHERE-1700-02-201304
53
Wavelength
range
400
nm –
500
nm
500
nm –
600
nm
600 nm –
700 nm
700 nm – 800
nm
800 nm – 900 nm
900 nm – 1100
nm
1.
Spec_Distr [%]
1.70
1.29
1.04
0.91
0.78
0.71
2.,1 Dark [%]
0.20 0.05
0.06
3.,1 T_TEC [%]
0.06
3.,2 StrayL [%]
ntia
Standard
Uncertainty [%]
1.71
1.29
1.04
0.91
0.78
0.71
Expanded
Uncertainty [%]
for k = 2
3.42
2.58
2.08
1.82
1.56
1.42
A missing value indicates that the value is less than 0.1 % or not known.
How to Estimate the Uncertainties for the Spectral Match
Test
An example illustrates the procedure for a measurement with a solar simulator. The following spectrum is
recorded.
RaySphere calculated the percentage of total irradiance in the wavelength range of 400 nm to1100 nm.