Lab 850
Operation
39
ba75553e03
01/2006
4.6.2
Resetting all meter settings
This function resets to the default condition all meter settings including
the calibration values. The corresponding values are given on the
following pages:
Resetting the meter
settings
Note
The measuring system is not calibrated after a reset. Before measuring
recalibrate the meter.
Calibration values
System settings
Measurement settings
1
Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2
During the display test, press <MODE> to open the menu for
the reset of the meter settings.
Init is displayed.
3
Press <
▲><▼> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.
4
Confirm with <OK>.
The menu is finished.
The meter switches to the measuring mode.