Calibration, Cond, Entry of solution” calibration – Xylem HandyLab 780 User Manual
Page 64: Cell constant calibration
Advertising

64
Cond
Calibration
“Entry of Solution” Calibration
(Calibration by entering the conductivity with display of the cell con-
stant)
1) Press the Start softkey.
2) Immerse the sensor in the solution.
3) Enter the temperature-corrected conductivity value and press enter.
4) Finally, you can Apply these values or Cancel the calibration.
Cell Constant Calibration
(Calibration by entering the cell constant with display of the conductivi-
ty)
1) Press the Start softkey.
2) Immerse the sensor in the solution.
3) Modify the value of the cell constant until the temperature-corrected
conductivity value is reached. Then press enter.
4) Finally, you can Apply these values or Cancel the calibration.
Note:
To abort calibration, you can press meas at any time.