Section 8 maintenance, 1 cleaning and storage, 2 probe precautions – YSI 30 User Manual

Page 29: 1 cleaning and storage 8.2 probe precautions

Advertising
background image

SECTION 8 MAINTENANCE

8.1 CLEANING AND STORAGE

The single most important requirement for accurate and reproducible results in conductivity
measurement is a clean cell. A dirty cell will change the conductivity of a solution by contaminating
it.

NOTE:

ALWAYS RINSE THE CONDUCTIVITY CELL WITH CLEAN WATER AFTER

EACH USE.

To clean the conductivity cell:

1. Dip the cell in cleaning solution and agitate for two to three minutes. Any one of the foaming

acid tile cleaners, such as Dow Chemical Bathroom Cleaner, will clean the cell adequately.
When a stronger cleaning preparation is required, use a solution of 1:1 isopropyl alcohol and

1 N HCl. Remove the cell from the cleaning solution.

2. Use the nylon brush (supplied) to dislodge any contaminants from inside the electrode chamber.

3. Repeat steps one and two until the cell is completely clean. Rinse the cell thoroughly in

deionized, or clean tap water.

4. Store the conductivity cell in the meter storage chamber.

8.2 PROBE

PRECAUTIONS

Observe the following precautions to ensure accurate, repeatable results:

1. The cell must be clean before making any measurements. When working with substances having

low conductivity, extraordinary cleanliness may be required.

2. Suspend the cell in the solution deep enough to submerge the vent hole. The electrode chamber

should be free of trapped air. You may need to tap the probe gently to dislodge any air bubbles.

3. Ideally, the cell should be at least ¼ inch away from any other object, including the sides or

bottom of the solution container.

YSI, Incorporated

Model 30/30M

25

Advertising
This manual is related to the following products: