Troubleshooting

Section

9

YSI, Incorporated

Model 30/30M

28

SYMPTOM

POSSIBLE

CAUSE

ACTION

10. Main Display reads “LErr”

A. In specific conductance mode,

temperature exceeds the values
computed using user defined
temperature coefficient and/or
reference temperature

B. During calibration, temperature

exceeds the values computed using
user defined temperature
coefficient and/or reference
temperature

A. & B. Adjust user defined

temperature coefficient and/or
reference temperature (See section
6 Advanced Setup)

11. Main Display reads “Err”

Secondary Display reads “ra”

A. System has failed its RAM test check

procedure

A. Turn instrument OFF and back ON.

B. Return the system for service.

12. Main Display reads “Err”

Secondary Display reads “ro”

A. System has failed its ROM test check

procedure

A. Turn instrument OFF and back ON.

B. Return the system for service.

13. Secondary Display reads “ovr”

A. Temperature is > 95

o

C

A. Read solution of lower temperature

B. Return system for service

C. Replace Probe/Cable assy.

14. Secondary Display reads “udr”

A. Temperature is < -5

o

C

A. Read solution of higher temperature

B. Return system for service

C. Replace Probe/Cable assy.

15. Secondary Display reads “rEr”

A. Temperature jumper is set to

o

F and

reading is >199.9

o

F but < 203

o

F

A. Set jumper to read

o

C.

B. Return system for service

The Model 30M has additional error codes as follows:

SYMPTOM

POSSIBLE

CAUSE

ACTION

1.

Readings on main display don’t
change

A. Meter is in recall mode.

A. Press MODE key to return to

Normal Operation

2.

Main Display reads “nOnE”

A. During recall, no data is currently

stored in memory.

A. Store data before attempting to recall

Advertising