YSI 30 User Manual

Page 30

Advertising
background image

Maintainence

Section

8

YSI, Incorporated

Model 30/30M

26

4. Stirring may be necessary for highest accuracy measurements, especially in low-conductivity

solutions.

5. If possible, isolate from ground potential the container or system in which measurements are to

be made.

6. Electrical fields and stray currents caused by stirrer motors, heaters, etc., can interfere with

measurements. The user should determine the effects of these and make the necessary
corrections, either by shielding or by disconnecting those units that cause trouble.

7. Always rinse the cell carefully before transferring it from one solution to another.

8. Never store a dirty or contaminated cell.

Advertising
This manual is related to the following products: