Trouble shooting guide, Symptoms, Rouble – YSI 5239 Dissolved Oxygen Probe User Manual

Page 16: Hooting, Uide

Advertising
background image

12

Stick the disc to the sanding tool, then wet sand the gold with
a twisting motion about 3 times or until all silver deposits are
removed and the gold appears to have a matte finish. If the
cathode remains tarnished, return the probe for service.

Wipe the gold cathode thoroughly with a wet paper towel
before putting on a new membrane cap.

T

ROUBLE

S

HOOTING

G

UIDE

Below is a quick reference of symptoms with steps to follow
to cure these symptoms.

Symptoms

Actions

Unable to calibrate

Extend the stabilization time from 15 min.
to 30 min. before calibrating

Change membrane

Check for loose or wrinkled membrane or
deposits between gold cathode and
membrane, especially after changing
membrane

Clean probe (see instructions under
Electrode Cleaning)