Symptoms, Servicing, Ervicing – YSI 5239 Dissolved Oxygen Probe User Manual

Page 17

Advertising
background image

13

Symptoms

Actions

Does not hold
calibration

•

Allow sufficient time for temperature and

oxygen stabilization before calibration

•

Check for water droplets on membrane

•

Check membrane for a pinhole or split

•

Change membrane

•

Clean probe (see instructions under
Electrode Cleaning)

Unstable readings

•

Check membrane for a pinhole or split

•

Allow 60 sec. for reading to stabilize

•

Change membrane

•

Clean probe (see instructions under
Electrode Cleaning)

Membrane cap is
tight

•

Use pliers to loosen the cap (turn counter
clockwise while facing sensor tip)

S

ERVICING

The 5239 probe has only two replaceable parts (other than the
membrane cap), the probe guard and the 5-pin MS connector
on the end of the cable. The cable is permanently attached to
the probe and cannot be replaced.

YSI part number

Description

110275

5-pin MS connector kit

059956

Probe guard

5238

Probe reconditioning kit