Stick the disc to the sanding tool, then wet sand the gold with
a twisting motion about 3 times or until all silver deposits are
removed and the gold appears to have a matte finish. If the
cathode remains tarnished, return the probe for service.

Wipe the gold cathode thoroughly with a wet paper towel
before putting on a new membrane cap.

T

ROUBLE

S

HOOTING

G

UIDE

Below is a quick reference of symptoms with steps to follow
to cure these symptoms.

Symptoms

Actions

Unable to calibrate

•

Extend the stabilization time from 15 min.
to 30 min. before calibrating

•

Change membrane

•

Check for loose or wrinkled membrane or
deposits between gold cathode and
membrane, especially after changing
membrane

•

Clean probe (see instructions under
Electrode Cleaning)