Symptoms, Servicing, Ervicing – YSI 5239 Dissolved Oxygen Probe User Manual

Page 17

Advertising
background image

13

Symptoms

Actions

Does not hold
calibration

Allow sufficient time for temperature and

oxygen stabilization before calibration

Check for water droplets on membrane

Check membrane for a pinhole or split

Change membrane

Clean probe (see instructions under
Electrode Cleaning)

Unstable readings

Check membrane for a pinhole or split

Allow 60 sec. for reading to stabilize

Change membrane

Clean probe (see instructions under
Electrode Cleaning)

Membrane cap is
tight

Use pliers to loosen the cap (turn counter
clockwise while facing sensor tip)

S

ERVICING

The 5239 probe has only two replaceable parts (other than the
membrane cap), the probe guard and the 5-pin MS connector
on the end of the cable. The cable is permanently attached to
the probe and cannot be replaced.

YSI part number

Description

110275

5-pin MS connector kit

059956

Probe guard

5238

Probe reconditioning kit