NitraVis 70x IQ TS

Commissioning

3 - 9

ba75950e02

11/2012

3.4.3 Settings for the main sensor

Measuring cycle

A measuring cycle consists of the cleaning procedure, the adjustment

time for the measuring system and the determination of the measured

value. The following graphic demonstrates the relevant settings:

Fig. 3-6 Measuring cycle of the UV-VIS sensor

Carrying out settings

Using <S>, switch from the measured value display to the main menu

of the settings. Then navigate to the setting menu (setting table) of the

sensor. The procedure is described in detail in your IQ S

ENSOR

N

ET

system operating manual.
Default values are marked in bold.

Time

Adjustment time

Measurement

Measuring

cycle

Measuring interval

Cleaning

Menu item

Settings

Explanations

Measuring mode

NO3-N

NO3

The measured parameter is displayed

in the selected citation form.

Measuring location

NitraVis 701 IQ TS

:

Activation

Inlet

Outlet

NitraVis 705 IQ TS

:

Outlet

(permanently set)

Measurement location or application of

the sensor.

The possible measurement loca-

tions are displayed depending on the

currently set measuring mode.

The sample matrix on which the cal-

culation of the measured value is

based changes with the location of

the sensor.

Measuring range

-

Display of the measuring range

(see chapter 8 T

ECHNICAL

DATA

).