Settings of the secondary sensor (tss measure, Ment) – YSI IQ S ENSOR N ET NitraVis 70x IQ TS User Manual

Page 21

Advertising
background image

NitraVis 70x IQ TS

Commissioning

3 - 11

ba75950e02

11/2012

3.4.4 Settings of the secondary sensor (TSS measurement)

Carrying out settings

Using <S>, switch from the measured value display to the main menu

of the settings. Then navigate to the setting menu (setting table) of the

sensor. The procedure is described in detail in your IQ S

ENSOR

N

ET

system operating manual.
Default values are marked in bold.

Cleaning type

Ultrasound

Air

Ultrasound+air

Cleaning methods

This menu item is only available with:

Cleaning mode

=

On

The duration of the cleaning procedure

is defined in the sensor for all cleaning

types.

Note: The duration of the cleaning pro-

cedure with air can be adjusted for the

linked relay in the menu,

150 - Settings

of outputs and links

/

Relay function

/

Sensorgesteuert

/

Pulsdauer

.

Please note the software requirements

for the MIQ/CR3, MIQ/R6, MIQ/CHVP-

lus, DIQ/S 182 relays (see section 3.1).

Save and quit

-

The system confirms the saving of the

settings and the display switches to the

next higher level.

Quit

-

The display switches to the next higher

level without saving the new settings.

Menu item

Settings

Explanations

Menu item

Settings

Explanations

Measuring mode

TSS

Display of the measuring mode

(TSS = total suspended solids measure-

ment)

Measuring range

-

Display of the measuring range

(see chapter 8 T

ECHNICAL

DATA

).

Cal - # raw value 1

Value pairs of the user calibration for TSS measurement (see section

4.3.2).

Note: Leave those values unchanged during the initial commission-

ing.

Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2