What to do if ...

6 - 1

ba76036e01

01/2012

6

What to do if ...

No compressed air at

the output

Cause

Remedy

– Compressed air supply

interrupted or too weak

– Check the compressed air

supply

– Supply voltage not present or

too low

IQ S

ENSOR

N

ET

:

– See the chapter, W

HAT

TO

DO

IF

... in the system manual

– Error in triggering

– Perform the function check

according to section 3.6.

– If the valve does not switch,

check the control

– loose clamping connection

– broken control line

– defective relay output

– valve output was not linked

with a sensor (see
section 4.2)

– incorrect setting of the linked

valve output (see chapter 4)