What to do if ...
No compressed air at
the output
Cause
Remedy
– Compressed air supply
interrupted or too weak
– Check the compressed air
supply
– Supply voltage not present or
too low
IQ S
ENSOR
N
ET
:
– See the chapter, W
HAT
TO
DO
IF
... in the system manual
– Error in triggering
– Perform the function check
according to section 3.6.
– If the valve does not switch,
check the control
– loose clamping connection
– broken control line
– defective relay output
– valve output was not linked
with a sensor (see
section 4.2)
– incorrect setting of the linked
valve output (see chapter 4)