Commissioning
ViSolid
®
700 IQ (SW)
3 - 8
ba76040e02
07/2013
3.4.2
Measurement in pipelines
Note
If there is a low level of total suspended solids (< 2000 mg/l SiO2 or 
< 1000 mg/l TSS), the effects of the measurement environment can 
simulate a higher content of total suspended solids. The effect of the 
measurement environment can be reduced by ensuring the optimum 
conditions (see section 3.3.1). 
Note
If an optimum installation is not possible due to the structural conditions 
at the measuring location (e.g. in narrow pipelines), the effects of the 
measurement environment can be compensated by user calibration 
(see section 4.2.4).
If deposits occur on the pipe walls, the calibration should be repeated 
at regular intervals.
Example:
45 ° pipe installation
The pipe should be straight for a length of approx. 25 cm beyond the 
installation location. Angled or tapered pipes can cause interference 
effects in the case of low levels of total suspended solids.
Fig. 3-5
Total suspended solids sensor in the pipe with EBST 700-DU/N flow-thru 
adapter
Fig. 3-5 shows the installation of the EBST 700-DU/N flow-thru adapter 
for installation in a pipeline (DN 50). The infrared beam points in the 
opposite direction to the direction of flow. The marking on the sensor 
points towards the pipeline (see Fig. 3-5).
Note
For exceptions to the direction of flow, see section 3.3.2 F
.
ADA-DF 9
EBST 700-DU/N
ca. 25 cm
straight pipe
Marking aid
in this direction
Swivel nut
Infrared beam
Direction
of flow