Xylem IQ SensorNet ViSolid 700 IQ (SW) User Manual
Page 21
 
ViSolid
®
700 IQ (SW)
Commissioning
3 - 9
ba76040e02
07/2013
Example:
90 ° pipe installation
Fig. 3-6
Total suspended solids sensor in a pipe (90 °)
The following points must be observed for a right-angled installation in 
the pipe (Fig. 3-6):
 Rotate the sensor so that the marking on the sensor points in the
direction of the pipe axis
 Select a position where the pipe diameter is as large as possible as
the installation location (see section 3.3.5 D
Note
In a 90° pipe installation with low levels of total suspended solids 
(< 2000 mg/l SiO2 or < 1000 mg/l TSS), the effects of the 
measurement environment can have a particularly significant effect on 
the measured value. Ensure the optimum conditions of the 
measurement environment (see section 3.3.1). 
Marking aid
Infrared beam
EBS 700-DU/N
Marking aid
in this direction
ADA-DF 9
1
Connect the SACIQ (SW) sensor connection cable to the plug 
head connector of the sensor and screw it tight (see section 
3.5.1).
2
Attach a marking aid (adhesive strips or similar) in the same 
position as the marking on the sensor to the plug head 
connector.