Commissioning, measuring, calibration

Commissioning

Example

4010-3 with

mounted 4010-2/3

pH Adapter DIN

Note
The further commissioning steps are given in the operating manual of the pH
electrode.

All further details are given in the operating manuals of the pH electrode and
meter.

1

Unscrew the 4 screws of the cover plate (the screw driver is included
in the scope of delivery).

2

Insert the 4010-2/3 pH Adapter in the open mounting slot.

3

Attach the adapter to the meter with the 4 screws.

1 pH electrode
2 Reference electrode
3 Temperature sensor (NTC 30)

4

Connect the electrode to the socket of the adapter.

5

Calibrate the electrode according to the operating manual of the me-
ter.

1

2 3