Technical data – YSI 4010-2 pH Adapter DIN / BNC User Manual

Page 5

Advertising
background image

4010-2/3 pH Adapter DIN / 4010-2/3 pH Adapter BNC

General information

5

ba76144e01

02/2013

Technical data

Note
The latest version of the present operating manual can be found on the Inter-
net under

www.YSI.com

.

General data

Electrodes that can be connected

All pH electrodes with DIN or BNC
plug head connector

Application

characteristics

pH measuring range

Depending on the electrode used

Allowed temperature range

0 ... 80 °C (32 ... 176 °F)

Typical application

Applications with special pH elec-
trodes

Measuring

characteristics of

the IDS measuring

electronics in the

4010-2/3 pH

Adapter

Measured pa-
rameter

Measuring range

Resolution

Accuracy
(± 1 digit)

pH

-2.000 ... 20.000

0.001

± 0.004

U [mV]

-1000.0 ... +1000.0

0.1

± 0.2

T [°C]

0.1

± 0.2

Material, electrical

connection

Material of the plug enclosure

ABS

Electrode connection

DIN or BNC

Meter connection

Mounting slot with contacts

Electrical data of

the electrode

connection

Input resistance

> 5 * 10

12

ohm

Input current

< 1 * 10

-12

A

Advertising
This manual is related to the following products: