Specifications, Accessories/spare parts, Recycling – YSI pH1200 User Manual

Page 9: Battery disposal, Specifications accessories/spare parts recycling

Advertising
background image

13

14

A C C E S S O R I E S / S P A R E P A R T S

Item

Description

601201

pH Electrode, Lab, 1M

601202

Temperature Probe, Stainless Steel

601203

pH/ATC Temp Combination Probe, Lab, 1M

605170

Electrode stand

603824

pH buffer, assorted case of US buffers. 2 pints ea of
buffer 4, 7 and 10

R E C Y C L I N G

YSI is committed to reducing the environmental footprint in the course
of doing business. Even though materials reduction is the ultimate
goal, we know there must be a concerted effort to responsibly deal
with materials after they’ve served a long, productive life-cycle. YSI’s
recycling program ensures that old equipment is processed in an
environmentally friendly way, reducing the amount of materials going
to landfills.

• Printed Circuit Boards are sent to facilities that process and reclaim

as much material for recycling as possible.

• Plastics enter a material recycling process and are not incinerated

or sent to landfills.

• Batteries are removed and sent to battery recyclers for dedicated

metals.

When the time comes for you to recycle, follow the easy steps outlined
at ysi.com.

BATTERY DISPOSAL

The 1200 is powered by alkaline batteries which the user must remove
and dispose of when the batteries no longer power the instrument.
Disposal requirements vary by country and region, and users are
expected to understand and follow the battery disposal requirements
for their specific locale.

S P E C I F I C A T I O N S

Display

Range

Accuracy

Resolution

pH

0.00 to 14.00

+/- 0.01 (instrument
only)

0.01 unit

Temperature 0.0 to 100°C

+/- 0.2°C

0.1°C

mV

-1999 to 1999
mV

+/- 0.05% +/- 1 digit 0.1 mV

Operating Temperature
Range

0 to 50°C (32 to 122°F)

Operating Relative
Humidity Range

up to 85%

Water resistance

IP-65 water resistant, splash resistant

Size

150 mm x 203 mm x 72 mm
5.9 in. x 8 in. x 2.8 in.

Weight with batteries

504 grams (1.1 lb)

Input impedance

>10

13

Ω

Power

Four 1.5 v C-size alkaline batteries
or 115/230V AC adapter

pH temp compensation

Auto/Manual 0.0 to 100.0°C

pH buffer recognition

Automatic recognition
US (4.01, 7.00, 10.01) or
NIST (4.00, 6.86, 9.18)

pH offset recognition

+/- 1.50 pH

pH slope recognition

30%

Connector (pH/temp)

BNC/8 pin DIN

Autolock feature

Yes

Backlit Display

Yes