6 orp, 1 measuring, 1 measuring the orp – YSI TruLab pH 1110 User Manual

Page 27: Measuring, Measuring the orp, 6orp

Advertising
background image

TruLab pH 1110

ORP

ba76155e01

03/2014

27

6

ORP

6.1

Measuring

6.1.1

Measuring the ORP

1.

Connect the ORP electrode to the meter.

2.

If necessary, call up the measured parameter U with <M>.

3.

Immerse the ORP electrode in the test sample.
The measured value is checked for stability (stability control).
The [AR] status display flashes.

4.

Wait for a stable measured value.
The [AR] display indicator no longer flashes.

Stability control (AutoRead )

During the measuring procedure, the stability control function is automatically
activated.

The stability control function (AutoRead) continually checks the stability of the
measured values in the monitored time interval. The stability has a consider-
able impact on the reproducibility of measured values. The [AR] display indi-
cator flashes until a stable value is measured.

Stability criteria (AutoRead )

3

6

1

8

2

4

U

°C

TP

mV

AR

3

Measured parameter

Time interval

Stability in the time interval

ORP

15 seconds

∆ : Better than 0.3 mV

Temperature

15 seconds

∆ : Better than 0.5 °C