TruLab pH 1110

pH value

ba76155e01

03/2014

21

8.

Wait for the measurement with stability control to be completed or termi-
nate the stability control with <ENTER>.
The calibration display for the next buffer appears.
Ct2 or Cd2 is displayed.

9.

If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.

or
Continue calibration using the next buffer with <ENTER>.

Continuing with two-point calibration

10. Thoroughly rinse the pH electrode with deionized water.
11. Immerse the pH electrode in the second buffer solution.
12. When measuring without temperature sensor: Enter the temperature of

the buffer with <><>.

13. Start the measurement with <ENTER> .

The measured value is checked for stability (stability control).
The [AR] status display flashes.
The electrode voltage (mV) or the nominal value of the buffer is displayed
(setting: see section 7.1.1 C

HANGING

THE

SETTINGS

FOR

P

H

MEASURE

-

MENTS

, page 29).

14. Wait for the measurement with stability control to be completed or termi-

nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears.
Ct3 or Cd3 is displayed.

For single-point calibration, the instrument uses the Nernst
slope (-59.2 mV/pH at 25 °C) and determines the zero point
of the pH electrode.