2 when to calibrate, 3 automatic calibration (autocal), When to calibrate – YSI TruLab pH 1110 User Manual

Page 20: Automatic calibration (autocal), 1t c

Advertising
background image

pH value

TruLab pH 1110

20

ba76155e01

03/2014

Calibration determines and stores the current values of the zero point and
slope. Thus, you should calibrate at regular intervals.

5.2.2

When to calibrate?

• When the calibration interval has expired
• Routinely within the framework of the company quality assurance
• After connecting another combination electrode

5.2.3

Automatic calibration (AutoCal)

Use any one to three buffer solutions of the selected buffer set in ascending or
descending order.

Below, calibration with YSI buffers (TEC) is described. When other buffer sets
are used, other nominal buffer values are displayed. Apart from that, the proce-
dure is identical.

1.

Start the calibration with <CAL>.
The calibration display for the first buffer appears.

2.

When measuring without temperature sensor: Temper the buffer or
measure the current temperature.

3.

If necessary, press <CAL> to select the buffer set used ([AutoCal TEC],
[AutoCal DIN]).
Ct1 or Cd1 is displayed.

4.

Thoroughly rinse the pH electrode with deionized water.

5.

Immerse the pH electrode in the first buffer solution.

6.

When measuring without temperature sensor: Enter the temperature of
the buffer with <><>.

7.

Start the measurement with <ENTER> .
The measured value is checked for stability (stability control).
The [AR] status display flashes.
The electrode voltage (mV) or the nominal value of the buffer is displayed
(setting: see section 7.1.1 C

HANGING

THE

SETTINGS

FOR

P

H

MEASURE

-

MENTS

, page 29).

8

4

2

1

T

C

°C

pH

AutoCal TEC

TP

Advertising