H-335 – Xylem H-335 User Manual
Page 29

H-335
SDI-12 Command and Response Protocol 2-13
Example of a H-335 Extended "Zero Offset" command:
Command
Response
Time Values
Description
"aXZO!"
"a0031<cr><lf>"
3 sec 1
Zero the UserAngle
Subsequent Command
Response
"aD0"
a+12.000<cr><lf>
The new UserAngleOffset
2.13 Extended “XZO” Command
This command is used to automatically zero the gate angle offset. The “aXZO” command
causes the H-335 to make a fresh angle measurement and use this value to automatically update the
UserAngleOffset value to produce a UserAngle reading = 0.00
Make certain the gate is fully closed and resting on the sill. Wait at least 15 minutes or so after
applying +12V power for the sensor temperature to stabilize before issuing the “aXZO” command.
This command should be used only during installation or maintenance.